AC Hipot test |
Rated Output | 200VA(5000V/40mA),short circuit current is greater than 200mA; |
Output Voltage | Range/accuracy | 100~5000V,±(1%of setting+2 counts) |
Output Frequency | Range/accuracy | 50Hz/60Hz,±0.1% of setting |
Alarm Current | Upper limit/ accuracy | 0.10~40.00mA,±(1%of setting+2 counts) |
Lower limit/ accuracy | 0.000~9.999mA,±(1%of setting+2 counts) |
Start Voltage | (0%~50%)× output voltage setting |
Current compensation | 0.000~40.00mA,Auto/Manual |
ARC detection | 1~9 level,9level is the most senstive,0 is off |
DC Hipot test |
Output Voltage | Range/accuracy | 100~6000Vdc,±(1%of setting+2 counts) |
Alarm Current | Upper limit/ accuracy | 0~10000μA,±(1%of setting+2 counts) |
Lower limit/ accuracy | 0.0~999.9μA,±(1%of setting+2 counts) |
Current compensation | 0 ~200.0μA,Auto/Manual |
Start Voltage | (0%~50%)×output voltage setting |
Charging Current lower Limit | (0~3.500)μA,Auto/Manual |
Discharge Time | ≤200ms |
ARC detection | 1~9 level,9 level is the most senstive,0 is off |
Insulation Resistance test |
Output Voltage | Range/accuracy | 100~2500Vdc,±(1%of setting+2V) |
Alarm Current | Upper limit/ accuracy | 100-499V,1~1000MΩ,±(8%of setting+2 counts) 1000~2000MΩ,±(12%of setting+2 counts) 500-2500V,1~199MΩ,±(2%of setting+2 counts), 200~999MΩ, ± (5%of setting+2 counts) 1000~9999MΩ, ± (15%of setting+2 counts) |
Lower limit/ accuracy |
|
|
|
|
Ground Bond test |
Rated Output | Max current 32A,Max test resistance:600mΩ,max open circuit voltage:12V |
Output Current | Range/accuracy | 2.0~32.0A,±(1%of setting+2 counts) |
Output Voltage | Range/accuracy | 3.0~6.4V ,±(1%of setting+2 counts) |
Output Frequency | Range/accuracy | Sin wave,50/60Hz,±0.1% of setting |
Alarm Limit | Resistance upper/lower limit | 2.0~10.6A:0.1~600mΩ; |
10.7~32.0A:0.1~R mΩ, R=(6400 / current setting)mΩ,As the current range is switched, the tester will auto calculate the max range |
Accuracy | ±(1%of setting+2 counts) |
Compensation Resistance | Range | 0~100 mΩ,automatic measurement, compensation ON/OFF |
Leakage Current test |
MD network selection | Standard network,IEC60990 fig4;Optional: other test network |
Probe setting | Leakage current to earth:G-L,G-N,AUTO(G-L,G-N),leakage current to frame:PH |
Power Supply status | The polarity conversion: open and close, automatic; The grounding and grounding transformation: on and off; The zero line of normal connection and unconnection transformation: on and off. |
|
|
Power supply | Single phase output, AN1640H need operate with optional isolated power supply,AN1651H with built-in 6kVA isolated power supply |
Upper/Lower Limit setting | Upper/Lower limit range: 0.0V~300.0V AC,resolution: 0.1V,45Hz~65Hz, |
Accuracy:±(0.4% of setting+0.1% of range) |
Max loading current | Max 20A |
Current:Upper/Lower limit (RMS) | Upper/Lower limit:0.0uA~999.9uA,1000uA~9999uA,10.00mA~12.00mA Resolution:0.1uA,1uA,0.01mA Accuracy: 0.0uA~999.9uA:DC:±(2% of setting +3 counts) 15Hz≤f≤100kHz:±(2% of setting +3 counts) 100kHz1000uA~7999uA:DC:±(2% of setting +3 counts) 15Hz≤f≤100kHz:±(2% of setting +3 counts) 100kHz8.00mA~12.00mA:DC:±(2% of setting+3 counts) 15Hz≤f<1000kHz,0.01mA~12.00mA:±5% of setting |
|
|
|
|
Current:Upper/Lower limit(peak value) | Upper/Low Limit:0.0uA~999.9uA,1000uA~9999uA,10.00mA~18.00mA Resolution:0.1uA,1uA,0.01mA Accuracy: 0.0uA~999.9uA:DC:±(2% of setting+2uA) 15Hz1000uA~7999uA:DC:±(2% setting+2 counts) 15Hz8.00mA~18.00mA:DC:±(2% of setting+2 counts) 15Hz≤f≤100kHz,0.01mA~18.00mA:±(10% of setting+2 counts) |
|
|
|
|
currentcompensation | Range:0.000~1.000mA,auto measure, ON or OFF |
|
Power test |
Power parameter | Alarm | Power upper/lower limit alarm; can turn on voltage, current , power factor upper/lower limit alarm |
|
Power upper/lower limit | Range:0.00W~99.99W,100.0W~999.9W,1000W~6000W Resolution:0.01W,0.1W,1W Accuracy:±(0.1% of setting+0.1%×range) |
|
|
Voltage upper/lower limit | When voltage upper/lower limit On: range:0.00V~99.99V,100.0V~300.0V,Resolution:0.01V,0.1V; Accuracy:±(0.1%of setting+0.1%×range),DC,45Hz≤f≤65Hz |
|
|
Current upper/lower limit | When current upper/lower limit On range:0.00mA~99.99mA,100.0mA~999.9mA,1.000A~9.999A,10.00A~20.00A Resolution:0.01mA,0.1mA,0.001A,0.01A Accuracy:±(0.1%of setting+0.1% of range),DC,45Hz≤f≤65Hz |
|
|
|
Power factor upper/lower limit | When power factor upper/lower limit On Range:±(0.100~1.000),Resolution:0.001 Accuracy:±0.01(Voltage/current amplitude is greater than the corresponding range of 10%) |
|
|
Frequency | Range:DC,45.00Hz≤f≤65.00Hz,Resolution:0.01Hz |
Voltage Start test |
Voltage Start test | Current upper/lower limit | range:(0.00~20.00)A,resolution0.01A Accuracy:±(0.1%of setting+0.1% of range) |
|
Current measurement | 20A,Crest Factor:1.6 range:(0.02~25.00)A,Resolution 0.01A Accuracy:±(0.1% of reading+0.1% of range), DC,45Hz≤f≤65Hz |
|
|
|
Voltage measurement | 300V,Crest Factor:1.6 Range:5.00V~99.99V,100.0V~300.0V,Resolution:0.01V,0.1V; Accuracy:±(0.1% of reading+0.1% of range),DC,45Hz≤f≤65Hz |
|
|
LN circuit detection |
LN circuit detection | Voltage | 23V AC/28V DC can choose, accuracy:±10% of reading |
Upper/lower setting | Range:(1.0~999.9)Ω,Resolution:0.1Ω,Accuracy:±(5%of setting+5 counts) |
Resistance measurement | Range:(1.0~999.9)Ω,Resolution:0.1Ω,Accuracy:±(5%×readinging+5 counts) |
Time |
Time | Ramp up time | 0.1~999.9s |
Ramp down time | 0,0.5~999.9s(0:infinite time) |
range | 0,0.5~999.9s(0:infinite time) |
Accuracy | ±(0.1%of setting+0.1s) |
General Specification |
Verification&Calibration | 23ºC±5ºC,(45~75)%RH ,less dust |
Operation | 0 ºC ~40ºC,,(5~90)%RH ,less dust |
Storage | -10ºC~55ºC,(5~95)%RH ,less dust |
Input Power | 220V±10%,50Hz±5% |
Power Consuption | Max 1000VA(excluding operation power of objects under test) |