Model | AN9651TH (F)(Three Phase/single phase7 in1) |
Ground Bond Test |
Empty Load Output Voltage | <12V |
Output Current Setting Range | AC3A~30A 50Hz/60Hz,1A/step |
Output Current Setting Accuracy | ±(2% of setting+0.2A) |
Upper Limit of Resistance | 3A≤output current≤10A, 10~600mΩ; 10A |
Ground Bond Resolution/ Measurement Range/ Accuracy | 1mΩ, 10~600mΩ, ±(2% of reading +2mΩ) |
Compensation Resistance Measurement Range/Accuracy | 0~200mΩ 1mΩ |
Test Time Setting Range | 0.5~999.9s 0.1s/step |
Insulation Resistance Test |
Rated Output Load/Current | 1VA (1000V/1mA) 1mA |
Output Voltage Range/Adjustable | DC:500V/1000V <1% (Empty load to full load) |
Output Voltage Ripple | Pure resistance load, 1000V/1mA, <3% |
Insulation Resistance Range | 1.0~2000MΩ |
Insulation Resistance Lower Limit Setting Range/ Resolution | 1.0~99.9 MΩ, 0.1MΩ/step, 100~2000 MΩ 1 MΩ/step |
Insulation Resistance Accuracy | 1~200 MΩ: ±(2% of reading+2counts) , 200~2000 MΩ: ±(5% of reading+5counts) |
Ramp-up Time Range/Resolution | 0.1~999.9s 0.1s |
Delay Judgment Time | 0.3~10.0s (Count from ramp-up ) |
Test Time Setting Range | 0.5~999.9s 0.1s/step |
AC Hipot Test |
Rated Output Load | 500VA (AC:5000V/100mA) |
Output Voltage Range /Frequency/Adjustable Rate | 500~5000V, 50Hz/60Hz, <2%(Empty load to full load) |
Output Voltage Setting Range/Accuracy | Cold state AC:500~5000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V) |
Leakage Current Preset Alarm | Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step |
Compensation Current Setting Range/ Resolution | 0.00 ~5.00mA, 0.01mA |
Output Initial Voltage Range | (0%~99%) of test voltage, from 0V to initial voltage need 200ms |
Leakage Current Measurement Range/Resolution/Accuracy | 0.10~99.99 mA, 0.01mA, ±(2% of reading +2counts) |
Test Time/Ramp-Up/Ramp-Down | 0.5~999.9s, 0.1~999.9s, 0.0~999.9s(After test pass) |
DC Hipot Test |
Rated Output Load | 60VA (AC:6000V/10mA) |
Output Voltage Range /Frequency/Adjustable Rate | 500~6000V, <2%(Empty load to full load) |
Output Voltage Setting Range/Accuracy | Cold state AC:500~6000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V) |
Leakage Current Preset Alarm | Upper limit: 50~9999μA; Lower limit: 0~9999μA , 1μA |
Compensation Current Setting Range/ Resolution | 0 ~5000μA, 1μA |
Output Initial Voltage Range | (0~99%) of test voltage, from 0V to initial voltage need 200ms |
Leakage Current Measurement Range/Resolution/Accuracy | 50~9999μA , 1μA, ±(2% of reading +2counts) |
Test Time/Ramp-Up/Ramp-Down | 0.5~999.9s, 0.1~999.9s, 0.3~10.0s(Count from ramp-up) |
Leakage Current Test |
Output Voltage Range | 1.06 times of input voltage (output by isolated transformer) |
Leakage Current Range/Resolution | Upper limit: 50~9999μA 1μA |
Compensation Current Setting Range/ Resolution | 0~500μA 1μA |
Test Time Range/Resolution | 1.0 ~999.9s 0.1s |
Leakage Type | dynamic, static |
Three Phase Dynamic Mode | A mode: ABC-AB-BC-CA (standard) B mode: ABC-AB-ABC-BC-ABC-CA (customized) |
Output Voltage Measurement Range/Accuracy | AC 50~500.0V 0.1V, ±(0.5% of reading+2counts) |
Leakage Current Measurement Range/Accuracy | 50~9999μA 1μA DC~10kHz: ±(2% of reading+5counts) 10kHz~1MHz: ±(5% of reading+5counts) |
Leakage Current Accuracy | 50.0~2000μA is ±(5% of reading);2000μA or above is ±(2.5% of reading) |
Leakage Power | 6kVA ,isolated transformer capacity is designed according to the rated voltage 220V |
Time Setting Range | 1~300s |
Measurement Mode | dynamic, static |
Power Test |
Output Voltage Range | 1.00 times of input voltage (output by isolated transformer) |
Power Setting Range/Resolution | Upper limit, Single phase:20~6000W Three phase:60~20000W Lower limit, Single phase:0~6000W Three phase:0~20000W 1W/step |
Test Time Range/Resolution | 0.5 ~999.9s 0.1s |
Voltage Measurement Range and Accuracy | AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts) |
Current Measurement Range and Accuracy | AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts) |
Active Power Measurement Range and Accuracy | 20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW 45~65Hz: ±(0.5% of reading +2counts) |
Power Factor Range and Accuracy | 0.100~1.000, ±(2% of reading +2counts) |
Low Voltage Start Test |
Output Voltage Range | 0.85 times of input voltage (output by isolated transformer) |
Test Time Range/Resolution | 0.5 ~999.9s 0.1s |
Current Setting Range/Resolution | Upper limit: 0.30 ~30.00A Lower limit: 0.00 ~30.00A0.01A |
Voltage Measurement Range and Accuracy | AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts) |
Current Measurement Range and Accuracy | AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts) |
Display Mode | 10.4 Inch, 640×480, TFT LCD display |
Isolated Transformer Capacity | Three Phase:20kVA, Single Phase: 6kVA |
Over Current Protection | Power, Start, Leakage Current, Phase Current35A, over 5s protection |
DUT Type Setting | Single Phase, Three phase Three line, Three phase Four line |
Grounding Mode | Load Grounding, Load Floating |
DUT Error Processing | Voice and light alarm |
Outer Alarm Light | Red alarm light |
PLC Output Interface | Each test function have one |
Interface Function | RS232, Outer Printer, Remote |
Optional Interface | RS485, GPIB, Internet Card, Inner Printer |
Remote control Interface | Foot Pedal, three-color alarm light(on-off signal output, PRINTER, 97 COM interface, Mini-Printer function optional) |
Test Group | 15 groups, 8 steps / group |
Buzzer Alarm Voice | Can turn on ,Can cut |