AC Hipot test |
Rated output capacity | AN9638H(F) | 500VA(5000V/100mA),short circuit current >200mA |
AN9637H(F) | 200VA(5000V/40mA),short circuit current >200mA |
Output voltage setting | Range,Accuracy | 100~5000V,±(2%×setting+5counts) |
Output frequency setting | Range,Accuracy | 50Hz or 60Hz,±0.1%×setting |
Alarm current limits setting | Upper-limit range,Accuracy | 0.10~100.00mA(or 40.00mA),±(2%×setting+5counts) |
Lower-limit range,Accuracy | 0.000~9.999mA,±(2%×setting+5counts) |
Time setting | Range | 0.5~999.9s(0 is infinite) |
Ramp up/down time range | 0.1~999.9s |
Accuracy | ±(0.2%×setting+1count) |
Measurement | Voltage range,Accuracy | 0.01~5.00kV,±(1.5%×reading+1count) |
Frequency range,Accuracy | 50Hz or 60Hz,±0.1%×reading |
Current range,Accuracy | 0.10~100.00mA(or 40.00mA),±(2%×reading+2counts) |
Time range,Accuracy | 0.1~999.9s,±(0.2%×reading+1count) |
Output voltage waveform, distortion, regulation | Sine wave,<2%(pure resistance load),<2%×setting+5V)(from empty load to full load) |
Start voltage setting | (0%~50%)×output voltage setting |
Current compensation setting | 0.000~100.00mA(or 40.00mA),auto,manual |
Arc detection | 1~9 level,9 level is optimally sensitive,0 is OFF |
DC Hipot test |
Output voltage setting | Range,Accuracy | 100~6000VDC,±(2%×setting+5counts) |
Alarm current limits setting | Upper-limit,Accuracy | 0~10000μA,±(2%×setting+5counts) |
lower-limit,Accuracy | 0.0~999.9μA,±(2%×setting+5counts) |
Time setting | Range | 0.5~999.9s(0 is infinite) |
Ramp up time range | 0.1~999.9s |
Ramp down time range | 0.3~999.9s(0 is OFF) |
Accuracy | ±(0.2%×setting+1 count) |
Measurement | Voltage range,Accuracy | 0.01~6.00VDC,±(1.5%×reading+1count) |
Frequency range,Accuracy | 0.0~10000uA,±(2%×reading+5count) |
Current range,Accuracy | 0.1~999.9s,±(0.2%×reading+1count) |
Rated output capacity | 60VA(6000VDC/10mA) |
Output voltage ripple, regulation | <5%(pure resistance load,6000V/10mA),<(2%×setting+5V)(from empty load to full load) |
Compensation current Setting | 0 ~200.0μA,auto, manual |
Start voltage setting | (0%~50%)× setting output voltage |
Charging current lower limit | (0~3.500)μA,auto, manual |
Discharge time | ≤200ms |
Arc detection | 1~9 level,9 level is optimally sensitive,0 is OFF |
Insulation Resistance test |
Output voltage setting | Range,Accuracy | 100~2500Vdc,±(2%×setting+5counts) |
Alarm resistance setting | Range | 1~9999MΩ |
Accuracy | 100-499V,1~1000MΩ,±(8%×setting+2counts) 1000~2000MΩ,±(12%×setting+2counts) 500-2500V,1~199MΩ,±(2%×setting+2counts), 200~999MΩ, ± (5%×setting+2counts) 1000~9999MΩ, ± (15%×setting+2counts) |
Time setting | Ramp up time range | 0.1~999.9s |
Adjustment delay time | 0,0.5~999.9s(0 is infinite) |
Test time range | 0,0.5~999.9s(0 is infinite) |
Ramp down time range | 0.1~999.9s |
Accuracy | ±(0.1%×setting+0.1s) |
Measurement | Voltage range, accuracy | 100~2500Vdc,±(2%×reading+5counts) |
Resistance range, accuracy | 1~9999MΩ; 100-499V,1~1000MΩ,±(8%×reading+2counts) 1000~2000MΩ,±(12%×reading+2counts) 500-2500V,1~199MΩ,±(2%×reading+2counts), 200~999MΩ, ± (5%×reading+2counts) 1000~9999MΩ, ± (15%×reading+2counts) |
Time range,accuracy | 0.1~999.9s,±(0.2%×reading+1counts) |
Charging current lower limit | 0~3.500μA,auto, manual |
Discharge time | ≤200ms |
Ground Bond test |
Rated output | AN9638H(F) | Max output current is 64A,max testing resistance 600mΩ,open-circuit voltage <12V |
AN9637H(F) | Max output current is 32A,max testing resistance 600mΩ,open-circuit voltage <12V |
Output current setting | Range,Accuracy | 2.0~32.0A(or 64A),±(2%×setting+5counts) |
Output voltage setting | Range,Accuracy | 3.0~10.0V ,±(2%×setting+5counts) |
Output frequency setting | Range,Accuracy | Sine wave,50Hz or 60Hz,±0.1%×setting |
Alarm limits setting | Resistance upper/lower limits | 2.0~10.6A:0.1~600mΩ; 10.7~32.0A(or 64A):0.1~R mΩ,among R=(6400 / set current value)mΩ,when current range changs,Tester for automatic calculation range scope biggest |
Tolerance | ±(2%×setting+5counts) |
Compensation resistance setting | Range | 0~100 mΩ,automatic measurement,ON/OFF |
Test time setting | Range,Accuracy | 0.5~999.9s(0 stand for time infinite length);±(0.2%×setting+1count) |
Measurement | Current range,accuracy | 2.0~32.0A(or 64A),±(2%×reading+2counts) |
Voltage range,accuracy | 3.0~10.0V ,±(2%×reading+5counts) |
Resistance range, accuracy | 10.0~99.9~100~600mΩ;±(2%×reading+5counts) (64A current calculated method is same) |
Time range,accuracy | 0.5~999.9s(0 stand for time infinite);±(0.2%×setting+1count) |
Extended function |
Communication function | Address setting | 0~999 |
Baud rate setting | 300,1200,2400,4800,9600,192000 |
Verification mode setting | None,odd,even |
Bar code scanning | Bar code scanning function | ON/OFF |
Bar code length setting | 1~30 |
Model identification number | Start bit select 1~30,max length 16 Bit |
Function allocation explain | Hardware interface is optional |
print function | printer type setting | stylus printer miniprinter |
| |